haven't had too many problems with the Alaris pumps, but the one thing I have noticed is that out of no where, the pump may pause itself, or the battery indicator light will start flashing (already plugged into power outlet) for no reason either. Have had our biomed dept look at those pumps when i find issues, and let them handle it. issues resolved after they inspect it.
We use alaris pumps and this has happened to me twice. A white screen with red bars at the top and bottom popped up along with a different-sounding beep and then the entire brain shut completely off stopping all four modules from infusing. I work ICU and both times vasopressors were on the pump. I reported it but don't know if they ever found out what caused it.
eftgirl
1 Post
Has anyone experienced Alaris Medley Guardrail IV pumps cutting off when bumped or sometimes from seemingly no touching at all? During spiking bags, walking patient, transporting, just moving pole, etc. Often with a 'connection' error. Sometimes there is an alarm, sometimes not.
Our pumps are about 5 years old. Starting out as uncommonly 'maybe someone turned it off' to now it's so widespread that we KNOW there is a problem and are afraid a patient will be injured. Just do not know the cause. FDA report submitted.
Company visited and is helping to discover the cause. However, they report only OUR hospital's pumps are having this widespread problem. We are investigating whether our cleaning products or other maintenance is the cause, but it seems unlikely our large medical center would be the ONLY ones having this problem. Looking at replacing connectors, although no guarantee this will fix the problem. Meanwhile how scary is it that an ICU patient's drips cut off??