FDA MedWatch - Class 1 Recall: Pulmonetic Systems LTV Series of Ventilators, Universal Cable Adaptor
MedWatch - The FDA Safety Information and Adverse Event Reporting Program
The Universal Cable Adaptor intended to correct an earlier Class I recall of
LTV series ventilators, Z-1485-04, is not functioning as intended. The
adaptor may not allow the ventilator to be powered up again if the
ventilator's internal battery has been depleted or may not be securely
attached to the pigtail connector on the ventilator. Customers have been
asked by the firm to remove the adaptors according to instructions provided
to them and return the adaptors to the company. Pulmonetic Systems'
permanent resolution of the ventilator's power switchover malfunction will
be to replace the power circuit board for all affected LTV Series
Ventilators.