You have to increase the pressure limit on the IV pump to overcome the pressure in the venous chamber. Some pumps can't. The arterial side has negative pressure so it isn't a problem if you run it on that side. Some nurses feel that more of the drug is removed if you pass it by the filter first but I disagree. Vancomycin is not protein bound and has a small molecular size. Running it on the venous side does not prevent the vanc from passing by the filter, it just delays it by a few minutes as it circulates through the blood stream and comes back through the filter.
reginalin, BSN, RN
28 Posts
Having problems whenever I administer Vancomycin. Whether it's pump (Braun Vista basic pump) or manual, the blood tries to back up the medication line. How to prevent the backing up because of the pressure in the venous chamber? Thanks.