Sounds like your attending was going for breaking the AFIB, not necessarily decreasing the heart rate. Although, I think Amiodarone is the first choice, I am thinking that in people who have chronic AFIB, the amiodarone is mostly ineffective, and you were probably right to suggest brevibloc, and at least stable him out by decreasing the heart rate. Honestly, amiodarone would have been my first choice as an anti arrhythmic, if it was new onset AFIB. Digoxin would have been my last choice, esp. immediately before heart surgery, w/ all the Potassium issues?
On your moments of terror subject, here is one I had happen a couple wks ago.
Didn't think it would ever happen, ever!? Well, all the stupid s%&t happened in one day. First case of the day was a TVH. checked the machine like I usually do, and everything was looking good, Pipeline pressure 50psi, and O2 cylinder 1800, good flow on flowmeters. 02 cylinder closed by me. CRNA rechecked to make sure everything was copacetic, saw the same reading on the O2 tank. I'm pretty sure he closed it, but did not recheck him.
About 10 minutes into the case after the patient is hooked up, pipeline pressure starts to fall, 30, 20, to zero. Alarm didn't go off until very late in the game, but it didn't matter. After wetting my pants, we switched to the O2 cylinder, and disconn. from wall. sorry, no pressure or gas there either. Okay, so I'll try medical air, but that's a no go too, cuz the machine must have some interlock that shuts down after the fail safe kicks in. Had to call stat for a new O2 tank, and anyone who knew what the heck was going on. Was baggin the patient manually through the machine temporarily on room air. never sufferred any deleterious consequences. sat. was 96 and above the whole time.
One of the connections mid line between the wall and the machine was cracked/loose a little bit. Everything appeared fine at the beginning, I think cuz everthing was fine. Then something freaky happened, and the connection came loose or something, and so the machine wasn't gettin any air so it went to the cylinder (which was supposedly closed), but maybe it was a faulty valve, and it wasn't shut tight all the way.
In any case, the pt. suffered no dire consequences, but even after we restored pipiline pressure, and put on a new O2 tank, the machine would alarm Reverse Flow, and Threshold Low. I know what they mean, but wasn't sure what it all added up to. Had our coordinator look at the machine. He adjusted the soda lime, said sometimes they need redistributed.
Any comments welcomed.